Instructions
Replace a faulty fuse. This problem will be indicated by the differential refractometer failing to power on, the display being blank or the fan refusing to engage. To replace, find the fuse holder slot, inches above the power input, and pull with minimum force to remove the holder from the rear panel. Place a new fuse in the same slot and turn on the machine.
Repair a gradually drifting baseline by changing the location of the unit and ensuring it is clear of air conditioning vents, chance breezes and direct sunlight.
Fix erratic retention times by cleaning, replacing or rebuilding (as required, once visually inspected) the pump check valve, as a malfunction with this component will cause this fault in the unit.
Visually inspect the type of cabling being used with the unit, and check that the wiring is properly shielded and in no way frayed or damaged. You should be using the shortest lengths of cabling possible, and the cables should be shielded. Making sure this is the case will fix excess electronic noise with the unit.